UNIQUE MEASUREMENT PROCEDURE WITH COMBICUFF.
| In the event that a patients limb circumference is borderline between |
| one bladder size and another |
| measurement with first one bladder size and then the other without |
| unduely disturbing the patient. |
| 1. Firstly, activate the smaller bladder by pressing down the smaller |
| size on the SELECTION VALVE then measure the patients blood |
| pressure and record the result. |
| 2. Secondly, activate the larger bladder by pressing down the larger size |
| on the selection valve and repeat and record the measurement. If it is |
| deemed necessary an average of the two results can be made. |